SPECIFICATIONS
|
|
AC WITHSTAND VOLTAGE
|
|
Range
|
Resolution
|
Accuracy
|
Output Voltage, ACV
|
0-5000
|
1
|
± (2% of setting + 5V)
|
Output Frequency
|
50Hz/60Hz ±0.1%, User Selection
|
Output Waveform
|
Sine Wave ,THD.<2% (Resistive Load), Crest Factor = 1.3 - 1.5
|
Output Regulation
|
± (1% of output + 5V), From no load to full load
|
SETTINGS
|
HI and LO-Limit (Total), current, mA
|
0.000-9.999 (0=OFF)
|
0.001
|
± (2% of setting + 2 counts)
|
10.00-40.00
|
0.01
|
HI and LO-Limit (Real), current, mA
|
0.000-9.999
|
0.001
|
± (3% of setting + 50µA)
|
10.00-40.00
|
0.01
|
Ramp-Up Timer, second
|
0.1 - 999.9
|
0.1
|
± (0.1% + 0.05 sec)
|
Ramp-Down Timer, second
|
0.0-999.9
|
Dwell Timer, second
|
0, 0.4 - 999.9
(0=continuous)
|
Ground Continuity
|
Current: DC 0.1 A ± 0.01A, fixed
|
Max. Ground Resistance: 1 Ω ± 0.1Ω, fixed
|
Arc Detection
|
1 - 9 ranges (9 is the most sensitivity)
|
DC WITHSTAND VOLTAGE
|
Output Voltage, DCV
|
0 -5000
|
1
|
± (2% of setting + 5V)
|
DC Output Ripple
|
<4 %(5KV/20mA at Resistive Load)
|
SETTINGS
|
HI and LO-Limit, current, µA
|
0.0 - 999.9
|
0.1
|
± (2% of setting + 2 counts)
|
1000 - 20000
|
1
|
Ramp-Up Timer, second
|
0.4 - 999.9
|
0.1
|
± (0.1% + 0.05 sec)
|
Ramp-Down Timer, second
|
0.0, 1.0-999.9
|
Dwell Timer, second
|
0, 0.3 - 999.9
(0=continuous)
|
Ramp-HI, current
|
>20 mA peak maximum, ON/OFF User Selection
|
Charge-LO, current
|
0.0 - 350.0 µA DC or Auto Set
|
Discharge Time
|
≦ 200 msec
|
Maximum Capacitive Load
DC Mode
|
1µF < 1KV 0.08µF < 4KV
|
0.75µF < 2KV 0.04µF < 5KV
|
0.5µF < 3KV
|
Ground Continuity
|
Current: DC 0.1 A ± 0.01A, fixed
|
Max. Ground Resistance: 1 Ω ± 0.1Ω, fixed
|
Arc Detection
|
1 - 9 ranges (9 is the most sensitivity)
|
INSULATION RESISTANCE
|
Output Voltage, DCV
|
50 - 1000
|
1
|
± (2% of setting + 2 counts)
|
Charging Current
|
Maximum >20mApeak
|
SETTINGS
|
HI and LO-Limit, resistance, MΩ
|
0.05 - 99.99
(HI – Limit: 0 = OFF)
|
0.01
|
Same as Resistance MEASUREMENT Accuracy
|
100.0-999.9
|
0.1
|
1000-50000
|
1
|
± (0.1% + 0.05 sec)
|
Ramp-Up Timer, second
|
0.1 – 999.9
|
0.1
|
Ramp-Down Timer, second
|
0=OFF, 1.0-999.9
|
Delay Timer, second
|
0, 1.0 - 999.9
(0=continuous)
|
Charge-LO, current, µA
|
0.000 - 3.500 or Auto Set
|
GROUND BOND
|
Output AC Current, A(7742 )
|
1.00 – 40.00
|
0.01
|
± (2 % of setting + 2 counts)
|
Output Voltage, Vac
|
3.00 – 8.00
|
0.01
|
± (2 % of setting + 3 counts)
|
Output Frequency, Hz
|
50Hz/60Hz ±0.1%, User Selection
|
Output Regulation
|
± (1% of output + 0.02A), Within maximum load limits, and over input voltage range.
|
Maximum Loading
|
1.00–10.00A/0–600mΩ, 10.01–30.00A/0–200mΩ, 30.01–40.00A/0–150mΩ
|
SETTINGS
|
Lead Resistance Offset, mΩ
|
0-200
|
1
|
1.00-2.99A,± (3 % of setting + 3 counts )
3.00 - 30A,± (2 % of setting + 2 counts )
3.00 - 40A,± (2 % of setting + 2 counts )
|
HI and LO-Limit Resistance, mΩ
|
0 – 150 (30.01-40.00A)
|
1
|
0 – 200 (10.01-30.00A)
|
0 – 600 (1-10.00A)
|
Dwell Timer, second
|
0, 0.5 - 999.9
(0 = continuous)
|
0.1
|
± (0.1% + 0.05 sec)
|
ConTINUITY TEST
|
Output Current, DC
|
0.1A ± 0.01A, fixed
|
SETTINGS
|
Lead Resistance Offset, Ω
|
0.00-2.00
|
0.01
|
± (3% of setting+2 counts)
|
HI and LO-Limit, Resistance, Ω
|
0.00 – 10.00 (0=OFF)
|
Dwell Timer, second
|
0,0.3-999.9
|
0.1
|
± (0.1% + 0.05 sec)
|
Opt.734 (0 - 10kΩ)
|
Output Current
|
DC 0.1A ± 0.01A,Max.
|
SETTINGS
|
Max-Lmt, Ω
|
0.00 – 9.99
|
0.01
|
± (1% of setting+5 counts)
|
10.0 – 99.9
|
0.1
|
± (1% of setting+2 counts)
|
100 – 999
|
1
|
1000 – 9999
|
1
|
Min-Lmt, Ω
|
0.00 – 9.99
|
0.01
|
± (1% of setting+5 counts)
|
10.0 – 99.9
|
0.1
|
± (1% of setting+2 counts)
|
100 – 999
|
1
|
1000 – 9999
|
1
|
Dwell Timer, second
|
0.0, 0.3 – 999.9
(0 = continuous)
|
0.1
|
± (0.1% + 0.05 sec)
|
Resistance Offset, Ω
|
0.00 – 2.00
|
0.01
|
± (1 % of reading + 5 counts)
|
MEASUREMENT
|
|
Range
|
Resolution
|
Accuracy
|
Voltage, KV(AC/DC)
|
0.00-5.00
|
0.01
|
± (1.5 % of reading +1 count)
|
Voltage, Vdc (IR only)
|
0-1000
|
1
|
± (1.5% of reading + 2 counts)
|
AC Current (Total), mA
|
0.000-3.500
|
0.001
|
± (2% of reading + 2 counts)
|
3.00-40.00
|
0.01
|
AC Current (Real), mA
|
0.000-9.999
|
0.001
|
± (3% of reading + 50?A)
|
10.00-40.00
|
0.01
|
All Ranges PF > 0.1 , V > 250VAC
|
DC Current, µA
|
0.0 - 350.0
|
0.1
|
± (2% of reading + 2 counts)
|
DC Current, mA
|
0.300 – 3.500
|
0.001
|
3.00-20.00
|
0.01
|
AC Current, A (G.B for 7742 only)
|
0.00-40.00
|
0.01
|
± (3 % of reading + 3 counts)
|
Resistance, MΩ (IR)
|
50-499V / 500-1000V
|
|
50 – 499V0.05–999.9, ±(7% of reading+2 counts)
|
0.050-1.999 / 0.050-9.999
|
0.001
|
500 – 1000V
|
2.00-19.99 / 0.050-9.999
|
0.01
|
0.10–999.9, ±(2% of reading+2 counts)
|
20.0-199.9 / 100.0-999.9
|
0.1
|
1000–9999, ±(5% of reading+2 counts)
|
200-50000 / 1000-50000
|
1
|
10000–50000, ±(15% of reading+2 counts)
|
Resistance, mΩ (GB)
|
0-600
|
1
|
1.00 – 2.99 A, ± (3 % of reading + 3 counts)
|
3.00 – 40.00 A, ± (2 % of reading + 2 counts)
|
Resistance, Ω (Continuity)
|
0.00-10.00
|
0.01
|
± (3 % of reading + 2 counts)
|
Resistance, Ω (Continuity) Opt.734
|
0.00 – 9.99
|
0.01
|
± (1% of reading+5 counts)
|
10.0 – 99.9
|
0.1
|
± (1% of reading+2 counts)
|
100 – 999
|
1
|
1000 – 9999
|
1
|
GENERAL
|
Input Voltage AC
|
115/230Vac±15%, 50/60Hz ± 5%, Fuse 10A/250V Slow-Blow for 7742 Model
|
PLC Remote Control
|
Input : Test, Reset, Interlock, Recall File 1 through 10
|
Output:- Pass, Fail, Test-in-Process
|
Memory
|
50 memories, 30 steps/memory
|
Graphic Display
|
320 x 240 Monographic LCD, 9 ranges contrast setting
|
Safety
|
Built-in Smart GFI circuit,GFI trip current 450µA max, HV shut down speed: <1mS
|
Interface
|
Standard RS232,Optional GPIB (IEEE-488.2),Multi-function Interface Card,Ethernet Card.
|
Security
|
Programmable password lockout capability to avoid unauthorized access to test set-up program.
|
Alarm Volume Setting
|
Range: 0-9 ;0=OFF, 1 is softest volume, 9 is loudest volume.
|
Calibration
|
Software and adjustments are made through the front panel. Automatic Calibration alert function to signal operator when re-calibration is due.
|
Environment
|
0-40℃, 20-80%RH
|
Dimensions/Net Weight
|
430mm(W)×133mm(H)×500 mm(D)/30Kg for MODEL 7742
|
STANDARD ACCESSORIES
|
Power Cord(10A)
|
×1
|
Fuses
|
×2 (Including a spare contained in the fuse holder)
|
Interlock Disable Key(1505)
|
×1
|
Hipot Test Lead ,1.5m(1101)
|
×1
|
Ground Bond Test Lead 40A
,1.6m(1137)
|
×1
|
Ground Bond Return Lead 40A
,1.6m(1138)
|
×1
|
RS232 link Cable 1.5m(1130)
|
×1
|
MATRIX SCANNER MODULE (Opt.736 for 7742)
|
High Voltage Rating
|
5KVAC/5KVDC
|
High Current Rating
|
40 A
|
Number of HV Channel
|
8
|
Number of HA Channel
|
8
|
RUN TEST MODULE (Opt.738,Opt.739 for 7742)
|
MEASUREMENT
|
|
Range
|
Resolution
|
Accuracy
|
Voltage, Vac
|
00.-277.0
|
0.1
|
± (1.5% of reading +2counts), 30.0-277.0V
|
Current, Aac
|
0.00-15.00
|
0.01
|
± (2.0% of reading + 2counts)
|
Power, Watts
|
0-4200
|
1
|
± (5% of reading + 3 counts)
|
Power, Factor
|
0.000-1.000
|
0.001
|
± (8% of reading + 2 counts)
|
Leakage Current, mA
|
0.00-10.00
|
0.01
|
± (2% of reading + 2 counts)
|
MD Circuit
|
Leakage Current measuring resistance = 2KΩ±1%
|
Timer, second
|
0.0-999.9
|
0.1
|
± (0.1% of reading + 0.05 sec)
|
SETTINGS
|
HI and LO Limit AC Voltage, V
|
0.0-277.0
|
0.1
|
± (1.5% of setting + 0.2 V), 30.0-277V
|
HI and LO Limit AC Current, A
|
0.00-15.00
|
0.01
|
± (2.0% of setting + 2 counts)
|
HI and LO Limit AC Power, W
|
0 - 4200
|
1
|
± (5.0% of setting + 3 counts)
|
HI and LO Limit Power Factor
|
0.000-1.000
|
0.001
|
± (8% of setting + 2 counts)
|
HI and LO Limit Leakage Current
|
0.00-10.00, Hi-Limit 0=OFF
|
0.01
|
± (2% of setting + 2 counts)
|
Delay Time, second
|
0.2-999.9
|
0.1
|
± (0.1% + 0.05 sec)
|
Dwell Time, second
|
0, 0.1-999.9 (0=continuous)
|
Cold Resistance
|
0~10KΩ
|
DUT POWER
|
AC Voltage
|
0 - 277.0V, Single phase unblance, 0 - 15.0A maximum
|
Power Rating
|
4200 W maximum
|
DUT Protection
|
Short circuit current 23A, <3 sec, Inrush Current 68A Response time 1m sec
|
STANDARD ACCESSORIES
|
Power Cord(1)
|
×1
|
LLT Receptacle Adaptor Box 20A ,1.6m(1905)
|
×1
|
LINE LEAKAGE TEST MODULE (Opt.739 for 7742)
|
CURRENT MEASUREMENT
|
Frequency Range
|
DC-1MHz
|
Leakage Current Range (RMS)
|
Resolution
|
Accuracy (DC-100KHz)
|
Accuracy ( >100KHz to 1 MHz)
|
0.0µA - 999.9µA
|
0.1µA
|
±(1.5% of reading + 3counts)
|
±(5% of reading + 5µA)
|
1000µA - 6000µA
|
1µA
|
±(1.5% of reading + 3counts)
|
Leakage Current Range (Peak), Opt.739 only
|
Resolution
|
Accuracy (DC-100KHz)
|
Accuracy ( >100KHz to 1 MHz)
|
0.0µA - 999.9µA
|
0.1µA
|
±(10% of reading + 2µA)
|
±(10% of reading + 2µA)
|
1000µA - 6000µA
|
1µA
|
LINE VOLTAGE MEASUREMENT
|
Range
|
0.0-277.0 VAC
|
Resolution
|
0.1 V
|
Accuracy
|
± (1.5% of reading +0.2V), 30.0-277.0V
|
DUT
|
DUT Input Power Rating
|
0-277V,AC@ 1ac max
|
DUT Protection
|
Short circuit current 23A, <3 sec, Inrush Current 68A Response time 1m sec
|
SETTINGS
|
High/Low Limit, µA
|
Range 0.0 - 6000µA(0=OFF) , Resolution 0.1µA
|
Delay Time, second
|
Range 0, 1.0-999.9 (0=continuous), Resolution 0.1second
|
Compliance Standards
|
A UL544 Non Patient, UL484, IEC60598, UL1363
|
B UL544 Patient Care
|
C UL2601-1, IEC60601-1, EN60601-1
|
D UL1563
|
E UL60950, IEC60950, IEC61010-1, IEC60335-1, IEC60990 Fig4 U2
|
F IEC60990 Fig5 U3
|
G IEC60990 Fig3 U1
|
MD A - G components
|
Accuracy: Resistance ± 1% Capacitor ± 5%
|
MD Voltage Limit
|
Maximum 30V peak or 30VDC
|
STANDARD ACCESSORIES
|
Hipot Return Lead ,1.8m(1102)
|
×2
|
*product specifications are subject to change without notice.
|