您当前的位置:首页 » 产品展示 » 直流稳压电源 » MX100T Aim-tti 三路输出稳压电源
产品名称: | MX100T Aim-tti 三路输出稳压电源 |
产品型号: | MX100T |
品牌: | 1532 |
产品数量: | |
产品单价: | 8800.00 |
日期: | 2024-09-19 |
MX100T Aim-tti 三路输出稳压电源的详细资料
MX100T and MX100TP:
OUTPUT 1 | |
---|---|
Ranges |
Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 16V at 1mA to 6A |
Operating Mode |
Co |
Voltage Setting | By direct numeric entry or quasi-analogue rotary wheel; resolution 1mV. |
Current Setting | By direct numeric entry or quasi-analogue rotary wheel; resolution 1mA. |
Setting Accuracy |
Voltage - 0.05% of reading ± 3mV; Current - 0.3% of reading ± 3mA to 3A, 0.5% of reading ± 3mA to 6A |
Setting Stores | Up to 50 set-ups can be saved and recalled via the keyboard (or the digital interfaces on MX100TP). |
Load regulation | <0.01% +5mV (CV mode) for any load change using remote sense. |
Line regulation | <0.01% +5mV (CV mode) for a 10% line voltage change. |
Ripple & Noise | Typically <0.5mV rms, <5mV pk-pk, 1mV rms max. (20MHz bandwidth). |
Transient Respo |
<100us to within 50mV of set level for 5% to 95% load change. |
Over Voltage Trip | Settable 1V to 40V in 0.1V steps |
Over Current Trip | Settable 0.1A to 7A in 0.01A steps |
Sensing |
Selec |
OUTPUT 2 | |
---|---|
Ranges |
Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 16V at 1mA to 6A Range 3 - 0V to 35V at 1mA to 6A* (available when output 3 is disabled) |
Operating Mode |
Co |
Voltage Setting | By direct numeric entry or quasi-analogue rotary wheel; resolution 10mV. |
Current Setting | By direct numeric entry or quasi-analogue rotary wheel; resolution 10mA. |
Setting Accuracy |
Voltage - 0.1% of reading ± 10mV; Current - 0.3% of reading ± 3mA to 3A, 0.5% of reading ± 3mA to 6A |
Setting Stores | Up to 50 set-ups can be saved and recalled via the keyboard (or the digital interfaces on MX100TP). |
Load regulation | <0.01% +5mV (CV mode) for any load change using remote sense. |
Line regulation | <0.01% +5mV (CV mode) for a 10% line voltage change. |
Ripple & Noise | Typically <0.5mV rms, <5mV pk-pk, 1mV rms max. (20MHz bandwidth). |
Transient Respo |
<100us to within 50mV of set level for 5% to 95% load change. |
Over Voltage Trip | Settable 1V to 40V in 0.1V steps |
Over Current Trip | Settable 0.1A to 7A in 0.01A steps |
Sensing | Selectable local or remote sensing. |
OUTPUT 3 | |
---|---|
Ranges |
Range 1 - 0V to 35V at 1mA to 3A; Range 2 - 0V to 70V at 1mA to 1. Range 3 - 0V to 70V at 1mA to 3A* (available when output 2 is disabled). |
Operating Mode |
Co |
Voltage Setting | By direct numeric entry or quasi-analogue rotary wheel; resolution 10mV. |
Current Setting | By direct numeric entry or quasi-analogue rotary wheel; resolution 1mA. |
Setting Accuracy |
Voltage - 0.1% of reading ± 10mV; Current - 0.3% of reading ± 3mA |
Setting Stores | Up to 50 set-ups can be saved and recalled via the keyboard (or the digital interfaces on MX100TP). |
Load regulation | <0.01% +5mV (CV mode) for any load change using remote sense. |
Line regulation | <0.01% +5mV (CV mode) for a 10% line voltage change. |
Ripple & Noise |
Typically <0.5mV rms, <5mV pk-pk, 1mV rms max. (20MHz bandwidth). 70V range - typically <1mV rms, <10mV pk-pk, 1.5mV rms max. |
Transient Respo |
<100us to within 50mV of set level for 5% to 95% load change. |
Over Voltage Trip | Settable 1V to 80V in 0.1V steps |
Over Current Trip | Settable 0.1A to 3. in 0.01A steps |
Sensing | Selectable local or remote sensing. |
Output Protection | |
---|---|
External Voltage | Output will withstand forward voltages of up to 50V (O/Ps 1 and 2) or 80V (O/P 3). Reverse protection by diode clamp, 3A max. |
Fault Trip |
The output will be shut down if a trip co |
Over Temperature |
Mo |
Sense Error |
Mo |
Co |
|
---|---|
Output Terminals |
Universal 4mm safety binding posts on 19mm (0·75”) spacing at front. Screw terminals at rear (MX100TP only). Terminals can accept fixed shroud 4mm plugs, standard 4mm plugs, fork terminals and bare wires. |
Sense Terminals | Sprung loaded screw-less terminals at front. Screw terminals at rear (MX100TP only). |
OUTPUT ON/OFF SWITCHING | |
---|---|
Individual On/Off | Individual keys for each output. On state indicated by key illumination. |
Multi-On/Multi-Off | Separate keys enable any combination of outputs to be turned on or off either simultaneously (default) or with timed delays from 10ms and 20s. Delayed operation indicated by flashing key illumination. |
VOLTAGE TRACKING | |
---|---|
The power supply can be set so that the voltage of an output is automatically set equal to that of another output and tracks any changes. | |
Tracking Modes | None, V2 tracks V1, V3 track V2, V2 and V3 both track V1. |
SETTING MEMORIES | |
---|---|
Individual Output Memories: | |
No. of Stores | 50 per output |
Parameters Stored | Range, Set Volts, Set Current, OVP, OCP |
l |
|
No. of Stores | 50 |
Parameters Stored | Range, Set Volts, Set Current, OVP, OCP (for all three outputs) |
METERING (each Output) | |
---|---|
Meter Function | 5 digit voltage and current meters (O/P 1), 4 digit voltage and current meters (O/Ps 2 & 3). Simultaneous display of actual and set values. |
Meter Resolution | 1mV/0.1mA (O/P 1), 10mV/1mA (O/Ps 2 & 3). |
Meter Accuracy | As per setting accuracy (CV mode). |
Additio |
V x A Calculated power in watts. Resolution 0.01Ohms. Accuracy 0·5% ± 3 digits |
DIGITAL BUS INTERFACES (MX100TP only) | |
---|---|
The MX100TP offers full remote co |
|
RS-232 | Standard 9-pin D connector. |
USB |
USB 2.0 co |
GPIB (IEEE-488) |
The interface co |
Ethernet (LAN) |
Standard 10/100 ba |
LXI Compliance | LAN interface is compliant with LXI core 2011. (LXI is the abbreviation for LAN eXtensions for Instrumentation). |
DIGITAL PROGRAMMING PERFORMANCE (MX100TP only) | ||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Programming Speed: | ||||||||||||||||||||||||||||||||||||
Command Delay | Typically <100ms between receiving the command terminator for a step voltage change at the instrument and the output beginning to change. | |||||||||||||||||||||||||||||||||||
Output Response: | ||||||||||||||||||||||||||||||||||||
|
DRIVER SOFTWARE SUPPLIED (MX100TP only) | |
---|---|
IVI Driver | An IVI driver for Windows is supplied. This provides support for common applications such as LabView*, LabWindows*, HPVEE* etc. |
USB Driver | An installation file is supplied which calls a standard Windows* USB driver. |
* LabView and LabWindows are trademarks of Natio * HPVEE (now Agilent VEE) is a trademark of Agilent Technologies. * USB interface is supported for Windows 2000, XP, Vista, 7 and 8 (including 64 bit versions). Windows is a trademark of Microsoft. |
GENERAL SPECIFICATIONS | |
---|---|
Input: | |
AC Input | 110V to 240V AC ±10%, 50/60Hz. Installation Category II. |
Input Power | 500VA max. |
Temperature & Environmental: | |
Operating Range | +5ºC to +40ºC, 20% to 80% RH |
Storage Range | -40ºC to + 70ºC |
Enviro |
Indoor use at altitudes up to 2000m, Pollution Degree 2. |
Cooling | Intelligent variable-speed fans. |
Safety & EMC: | |
Safety | Complies with EN61010-1 |
EMC | Complies with EN61326 |
Physical: | |
Size | 212 x 130 x 375mm (WxHxD) (half rack x 3U height) |
Weight | 4.8kg (MX100T); 4.9kg (MX100TP). |
OPTIONS: | |
Rack Mount | 19 inch rack mount for one or two power supplies. |